Abstract

Lays down detailed specifications for the instrumentation and procedure to be used for primary calibration of accelerometers at low frequencies. Applies to all kinds of pick-ups. Is applicable for a frequency range from 0,5 Hz to 20 Hz and a dynamic range from 1 m/s^2 to 200 m/s^2 (frequency-dependent) and from 0,5 mm to 1000 mm. The limits of uncertainty applicable are ± 0,5 % of reading.


General information 

  •  : Withdrawn
     : 1993-12
  •  : 1
     : 6
  •  : ISO/TC 108 Mechanical vibration, shock and condition monitoring
  •  :
    17.160 Vibrations, shock and vibration measurements